Menu
GHP 900
Conforming to the guarded hot plate standard ISO 8302, ASTM C177, EN 1946-2, EN 12664, EN 12667, and EN 12939, the GHP 900 series thermal conductivity testers are specifically designed to excel with thicker, high-density specimens. Their versatility extends to a broad range of applications. A unique feature of the GHP 900 series is the ability to introduce samples into the test chamber from any orientation. This facilitates the accommodation of bulky and intricate samples without compromising the integrity of the measuring plates. Data acquisition and instrument control are managed by an external Lambda Control desktop unit in conjunction with a Windows PC and Lambda software.
Specifications
Specifications
Measuring range | 0.005 to 2.0 W/(m·K), depending on material and thickness |
Specimen size (L x W) | 900 mm x 900 mm
variable, according to the dimension of the hot plate: 200 mm x 200 mm up to 500 mm x 500 mm |
Specimen thickness (H) | 1x 15 mm to 280 mm (1-specimen measurement method)
2x 15 mm to 140 mm (2-specimen measurement method) |
Temperature range | Cooling plate: -10°C to 60°C
Heating plate: 0°C to 70°C |
Interface 1 | 1x RS 232,1x Gigabit Ethernet |
Dimensions (H x W x D) | 233 cm x 146 cm x 126 cm |
Power supply | 110 V – 230 V, 50/60 Hz |
Weight | 278 kg |